Abstract

This innovative 3D phantom allowed the acquisition of reproducible nephrograms with parameters comparable to those of clinical examinations. At low activity, 3D-CZT camera acquisitions provided equivalent values to conventional cameras, supporting their use for qualitative imaging and follow-up.

Keywords

3D imaging3D-CZTRenal phantomSPECT[99mTc]-MAG3

Affiliated Institutions

Related Publications

Publication Info

Year
2025
Type
article
Citations
0
Access
Closed

Social Impact

Social media, news, blog, policy document mentions

Citation Metrics

0
OpenAlex
0
Influential

Cite This

Gilles Metrard, Gilles Le Rouzic, Benoit Chapelle et al. (2025). Renal dynamic scintigraphy using 3D-ring CZT cameras: comparison with NaI(Tl). EJNMMI Research . https://doi.org/10.1186/s13550-025-01351-2

Identifiers

DOI
10.1186/s13550-025-01351-2
PMID
41369969

Data Quality

Data completeness: 77%